Testing device Testor 2500 Draeger/R62950
**5.5 kg (12.1 lbs)**
## English Specification
1. **Application**
Manual static tightness / leak test for full face masks, SCBA, lung demand valves, chemical protective suits; compliant with VFDB 0804 guidelineDräger Saf…
2. **Compressed Air Supply**
Supply pressure: 4 … 10 bar (dry compressed air)
Built-in pressure reducer, independent of inlet pressure fluctuation
3. **Pressure Gauges**
– Medium-pressure gauge: 0 … 10 bar, scale division 0.5 bar, accuracy ±1.6% of full scale
– Low-pressure gauge: -15 / 0 / +25 mbar, scale division 0.5 mbar, accuracy ±1.0% of full scale
Supports both negative pressure and positive pressure test
4. **Timer**
LCD timer, adjustable range: 1 sec … 99 min, preset test duration for leak hold test
5. **Mechanical & Dimension**
Dimension (D × W × H): 300 × 515 × 335 mm
Integrated test head (Gel face II simulation face)
Bench-top installation, manual lever control panel
No PC software required (Testor 3500 is computer-controlled version)
6. **Ambient Operating Conditions**
Operating temperature: +10 ℃ … +40 ℃
Ambient barometric pressure: 850 mbar … 1400 mbar
Relative humidity: max. 70 %, non-condensing